Universal low-frequency (from 20 kHz to 2,5MHz) flaw detector is intended for detection defects (discontinuity and uniformity of materials) in half-finished products and finished articles, for measurement depth and their coordinates position, measurements a thickness, measurement velocity propagation and attenuation of ultrasonic oscillations in composite materials, plastic, concrete and in materials with the big attenuation. A pulse of excitation in the form a radio-frequency pulse with amplitude 200V both regulated frequency and quantity of the periods, and also the high-sensitivity amplifier and possibility to connect the preliminary amplifier, allow to realize the testing with application noncontacts (working through air) transducers. Completely a digital path with processing a radio signal and interface RS232 for personal computer connection, allow to realize various techniques of the testing of products - with processing not only amplitudes of a signal, but also a frequency spectrum or phase components of a signal.
 
| Range: up 160 to 10000 mks  Display: LCD or ELD 110 x 65 mm (240 х 128 pixels) with adjustable contrast and backlight  Amplifier: 20 кHz to 2500 kHz;  Controllable gain: 90 dB with step 0.5, 1, 2 and 6 dB  Probe impulse: a radio impulse amplitude 200 V, with established frequency of filling and number of the periods  Pulse repetition frequency: 10, 25, 100, 200 and 400 Hz  Rectification: positive halfwave, negative halfwave, full-wave, radiosignal (in all ranges)  Reject: linear, up to 0 to 80% screen height  | Monitor gate: two independent monitor gates with individual logic, sound indication  "Freeze" А-signal  Measurement of a thickness on one signal or between two signals  Memory of options: 64 adjustments about possibility of input of the name  Memory of results: 10 files on 100 results of measurement  Communication with PC: interface RS232  Mains power: 4 batteries size "D" or external block  8 hours at the disconnected backlight  Weight: 3 kg with batteries  | 
 
Basic Delivery Set 
UD2Н-PМ main unit 
Power supply 220 V 
Probe cable SR50-SR50 2 pieces. 
Transducers 60 кHz 2 pieces. 
Cable for connection PC 
Software support 
Documentations 
Transport bag 
Additional implements
External preliminary amplifier 
Batteries "D" 
Complete set of batteries the size "D" 9000 mA/hr (4 pieces.) 
Universal recharger 
Protective bag 
Scanner with wheel transducers 
Scanner with noncontact transducers 
Ultrasonic transducers for diverse application 
Technical Characteristics 
Calibration Range 
min.: 0 - 160 mks
max.: 0 - 10000 mks 
Sound velocity range 
100 - 4000 m/s 
Memory 
64 datasets; 10 files on 100 results 
Pulse shift 
up 0 mks to 7350 mks 
Pulse intensity 
amplitude 200 V 
Amplifier 
broadband from 20 to 2500 кHz with filters on 50, 
100, 200 and 400 kHz 
Controllable gain 
90 dB, with step 0.5, 1, 2 or 6 dB 
Rectification 
positive or negative half wave, complete, 
radiofrequency signal 
Reject 
linear, 0 - 80% screen height 
Monitoring gate 
two intependent monitor gates with individual logic,
sound indication 
Sound path measurement 
up to 0 to first signal in zone or between signals in zones 
Measurement resolution 
1 mks in all range 
Display of amplitude 
as a procentage of the screen height 
Display 
LCD with adjustable contrast and backlight,
ELD with adjustable brightness 
110 х 65 mm; 240 Х 128 pixels 
А-signal 
70 mm x 50 mm; 160 x 100 pixels 
Interface 
RS232 
Dimensions transducer 
2 SR50 
Batteries 
4 NiMh size "D" 
Operating time from batteries 
LCD: 10-14 hours  (depending on installations) 
ELD: 6-8 hours  (depending on installations)
Mains power 
via external plug-in power unit 220 V AC 
Operating voltage 
7.2 - 10 V DC 
Power consumption 
maximum 5 W, depending on 
instrument setting 
Operating temperature 
LCD: up to -5°C to +50°C 
ELD: up to -25°С to +55°С 
Dimensions (H x W x D) 
220 mm x 120 mm x 150 mm 
Weight 
2,5 kg with batteries